FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1955121 · Received January 10, 2011

Report

Report Number
1423500-2011-00369
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
December 16, 2010
Report Date
December 16, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 WAS NOT CONFIRMED DUE TO THE LACK OF A SAMPLE; HOWEVER, BASED ON THE INFORMATION GATHERED DURING BAXTER'S INVESTIGATION, THE CAUSE OF THE SYSTEM ERROR 2240 WAS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER A SUPPLY BAG FELL AND DISCONNECTED. THE LOT INFORMATION WAS UNKNOWN; THEREFORE A BATCH REVIEW WAS NOT PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS DISCARDED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON THE COMPLETION OF BAXTER'S INVESTIGATION.

Description of Event or Problem · 1

A PATIENT CONTACTED GLOBAL TECHNICAL SERVICES (GTS) REGARDING ASSISTANCE WITH A SYSTEM ERROR 2240 WHILE USING THE HOMECHOICE (HC) DURING THERAPY. THE PATIENT STATED THE ERROR WOKE HIM UP, AND HE NOTICED THE SUPPLY BAG HAD DISCONNECTED. GTS HAD THE PATIENT CYCLE POWER AND EXPLAINED THEY WOULD NEED TO START OVER WITH NEW SUPPLIES. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 60 YR HOMECHOICE APD CYCLER (B)(4) (REFURBISHED)