AUTOMATED PD SET W/CASSETTE4 PRONG
Report
- Report Number
- 1423500-2011-00369
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- December 16, 2010
- Report Date
- December 16, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 WAS NOT CONFIRMED DUE TO THE LACK OF A SAMPLE; HOWEVER, BASED ON THE INFORMATION GATHERED DURING BAXTER'S INVESTIGATION, THE CAUSE OF THE SYSTEM ERROR 2240 WAS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER A SUPPLY BAG FELL AND DISCONNECTED. THE LOT INFORMATION WAS UNKNOWN; THEREFORE A BATCH REVIEW WAS NOT PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4).
(B)(4). THE SAMPLE WAS DISCARDED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON THE COMPLETION OF BAXTER'S INVESTIGATION.
A PATIENT CONTACTED GLOBAL TECHNICAL SERVICES (GTS) REGARDING ASSISTANCE WITH A SYSTEM ERROR 2240 WHILE USING THE HOMECHOICE (HC) DURING THERAPY. THE PATIENT STATED THE ERROR WOKE HIM UP, AND HE NOTICED THE SUPPLY BAG HAD DISCONNECTED. GTS HAD THE PATIENT CYCLE POWER AND EXPLAINED THEY WOULD NEED TO START OVER WITH NEW SUPPLIES. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOMATED PD SET W/CASSETTE4 PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | HOMECHOICE APD CYCLER (B)(4) (REFURBISHED) |