FDA Adverse Event Injury Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 1955103 · Received January 10, 2011

Report

Report Number
2017865-2011-00067
Event Type
Injury
Date Received
January 10, 2011
Date of Event
November 12, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED ICD ANALYSIS FOUND AN ARC MARK ON CAN CHARACTERISTIC OF A DAMAGED LEAD.

Description of Event or Problem · 1

IT WAS REPORTED THAT NOISE WAS OBSERVED ON THE EGM. AN ALERT MESSAGE FOR POSSIBLE OUTPUT CIRCUIT DAMAGE WAS OBSERVED. AN INSULATION ANOMALY WAS SUSPECTED. THE LEAD WAS CAPPED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1580/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention