FDA Adverse Event Malfunction Summary report: N

RIATA PASSIVE FIXATION

MDR report key: 1955084 · Received January 10, 2011

Report

Report Number
2017865-2011-00022
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
November 4, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

PATIENT PRESENTED TO THE CLINIC AFTER RECEIVING INAPPROPRIATE SHOCK DUE TO OVERSENSING. X-RAY REVEALED A SUSPECTED INSULATION ANOMALY. THE PHYSICIAN TURNED OFF THE TACHY THERAPY. SYSTEM REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA PASSIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1570/65 NA

Patients

Seq Age Sex Outcome Treatment
1