FDA Adverse Event Injury Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 1955020 · Received January 10, 2011

Report

Report Number
2017865-2011-00078
Event Type
Injury
Date Received
January 10, 2011
Date of Event
November 8, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT NOISE, LOW IMPEDANCE, AND HIGH CAPTURE THRESHOLDS WERE OBSERVED. WHEN THE POCKET WAS OPENED, A SMALL INSULATION BREAK WAS OBSERVED AT 3CM FROM THE IS-1 CONNECTOR. THE LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1580/65 NA

Patients

Seq Age Sex Outcome Treatment
1 84 YR Required Intervention