FDA Adverse Event
Malfunction
Summary report: N
RIATA ACTIVE FIXATION
MDR report key: 1955012
·
Received January 10, 2011
Report
- Report Number
- 2017865-2011-00058
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- October 29, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT RECEIVED AN ALERT STATING THAT THE HIGH VOLTAGE LEAD IMPEDANCE FOR THE SVC-CAN VECTOR HAD INCREASED. RECOMMENDED A LEAD INTEGRITY TEST WHILE PERFORMING THE HVLI TESTING. THE PHYSICIAN OPTED TO MONITOR THE PATIENT. THE DEVICE AND LEAD REMAIN IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1580/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |