FDA Adverse Event Injury Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 1954986 · Received January 10, 2011

Report

Report Number
2017865-2011-00091
Event Type
Injury
Date Received
January 10, 2011
Date of Event
October 22, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DFT TESTING THE DEVICE FAILED TO DELIVER A SHOCK. AN ALERT OF POSSIBLE OUTPUT CIRCUIT DAMAGE WAS OBSERVED. THE PATIENT WAS EXTERNALLY RESCUED. AN INSULATION BREAK WAS OBSERVED DURING THE EXPLANT PROCEDURE. THE LEAD WAS CAPPED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1581/65 NA

Patients

Seq Age Sex Outcome Treatment
1 45 YR Required Intervention