FDA Adverse Event
Injury
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
MDR report key: 1954930
·
Received January 10, 2011
Report
- Report Number
- 2017865-2011-00278
- Event Type
- Injury
- Date Received
- January 10, 2011
- Date of Event
- October 14, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ANALYSIS REVEALED THAT THE IS-1 PROXIMAL CABLE WAS FRACTURED AT 4.0CM FROM THE CONNECTOR PIN. THIS COULD CAUSE THE REPORTED THRESHOLD AND IMPEDANCE MEASUREMENT ANOMALIES.
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Description of Event or Problem · 1
IT WAS REPORTED THAT UPON INTERROGATION OF THE DEVICE, ELEVATED CAPTURE THRESHOLD AND IMPEDANCE WERE OBSERVED. A LEAD FRACTURE WAS NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7120Q/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Required Intervention |