FDA Adverse Event Malfunction Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 1954869 · Received January 10, 2011

Report

Report Number
2017865-2011-00241
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
October 29, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4). FAILURE(EVENT) OBSERVED DURING ANALYSIS. ANALYSIS FOUND MEDICAL ADHESIVE ON THE HELIX PREVENTING THE HELIX FROM RETRACTING. THIS IS A WORKMANSHIP ANOMALY.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120/65 NA

Patients

Seq Age Sex Outcome Treatment
1