FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 1954856 · Received January 10, 2011

Report

Report Number
2017865-2011-00232
Event Type
Injury
Date Received
January 10, 2011
Date of Event
October 29, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED A PARTIAL LEAD WAS RETURNED, 12CM FROM THE CONNECTOR PIN. NO ANOMALY WAS DETECTED ON THE PARTIAL LEAD.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED TO A FOLLOW-UP AFTER RECEIVING INAPPROPRIATE THERAPY DUE TO NOISE. THE HIGH VOLTAGE LEAD IMPEDANCE HAD INCREASED. THE LEAD WAS CAPPED AND A PARTIAL LEAD WILL BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention