FDA Adverse Event Injury Summary report: N

CURRENT PLUS VR, DF-4 CONNECTOR

MDR report key: 1954795 · Received January 10, 2011

Report

Report Number
2017865-2011-00458
Event Type
Injury
Date Received
January 10, 2011
Date of Event
November 1, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
P910023
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. .OTHER TEXT : DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

THE REPORTED PREMATURE BATTERY DEPLETION WAS VERIFIED IN THE LABORATORY. BASED ON ALL AVAILABLE PARAMETER AND USAGE INFORMATION, THE DEVICE LONGEVITY WAS FOUND TO BE BELOW THE EXPECTED LIMITS. NO HIGH CURRENT DRAIN SOURCES WERE FOUND DURING BENCH, AUTOMATED ELECTRICAL, TEMPERATURE, AND HUMIDITY TESTING. THE BATTERY WAS SENT OUT TO THE VENDOR FOR FURTHER EVALUATION. AN INTERNAL BATTERY ANOMALY WAS FOUND TO CAUSE THE PREMATURE BATTERY DEPLETION.

Description of Event or Problem · 1

PATIENT PRESENTED TO THE CLINIC WITH THE DEVICE AT ERI. PATIENT REPORTED RECEIVING A VIBRATORY ALERT. THE DEVICE WAS EXPLANTED DUE TO SUSPECTED PREMATURE BATTERY DEPLETION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CURRENT PLUS VR, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD1211-36Q NA

Patients

Seq Age Sex Outcome Treatment
1 86 YR Required Intervention