FDA Adverse Event
Malfunction
Summary report: N
CURRENT PLUS DR, DF-4 CONNECTOR
MDR report key: 1954752
·
Received January 10, 2011
Report
- Report Number
- 2017865-2011-00474
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- October 4, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. RELIABILITY LABORATORY TECHNICIAN (B)(4) - FAILURE (EVENT) OBSERVED DURING ANALYSIS FOUND THE DEVICE WITH PREMATURE BATTERY DEPLETION. THE DEVICE WAS ANALYZED WITH A NEW BATTERY AND FOUND TO BE NORMAL. THE ORIGINAL BATTERY WAS SENT OUT TO THE VENDOR FOR FURTHER EVALUATION, AND NO ANOMALY WAS DETECTED THAT WOULD CAUSE THE BATTERY TO DEPLETE. THE CAUSE FOR THE PREMATURE BATTERY DEPLETION COULD NOT BE DETERMINED.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CURRENT PLUS DR, DF-4 CONNECTOR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | CD2211-36Q | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |