FDA Adverse Event
Injury
Summary report: N
ATLAS II VR
MDR report key: 1954680
·
Received January 10, 2011
Report
- Report Number
- 2017865-2011-00548
- Event Type
- Injury
- Date Received
- January 10, 2011
- Date of Event
- June 22, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED LATE DUE TO DELAY IN RECEIVING PAPERWORK. UPON INTERROGATION, NO COMMUNICATION COULD BE ESTABLISHED WITH THE DEVICE DUE TO A DEPLETED BATTERY. THE DEVICE WAS ANALYZED WITH A POWER SUPPLY AND WAS FOUND TO BE NORMAL. THE BATTERY WAS SENT OUT TO THE VENDOR FOR FURTHER EVALUATION, AND NO ANOMALY WAS DETECTED THAT WOULD CAUSE THE BATTERY TO DEPLETE. THE CAUSE FOR THE PREMATURE BATTERY DEPLETION COULD NOT BE DETERMINED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED DUE TO PREMATURE BATTERY DEPLETION TO EOL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATLAS II VR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | V-168 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Required Intervention |