FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 1954549 · Received December 15, 2010

Report

Report Number
1824206-2010-11615
Event Type
Malfunction
Date Received
December 15, 2010
Date of Event
November 19, 2010
Report Date
November 19, 2010
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH TIGHTENED ALL OF THE GROUND SCREWS THAT SECURE GROUND WIRES TO THE BED. THE BED PASSED THE ELECTRICAL SAFETY TEST AFTER HE TIGHTENED THE SCREWS.

Description of Event or Problem · 1

THE TECH FOUND THAT THE GROUND RESISTANCE WAS TOO HIGH ON THE BED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY BED BIRTHING BED HDD HILL-ROM, INC. 3600

Patients

Seq Age Sex Outcome Treatment
1