FDA Adverse Event Malfunction Summary report: N

REALIZE ADJ GASTRIC BAND STR

MDR report key: 1954517 · Received January 10, 2011

Report

Report Number
3005992282-2011-00010
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
December 17, 2010
Report Date
November 30, 2010
Manufacturer
OBTECH MEDICAL SARL_
Product Code
LTI
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INJECTION PORT WITH THE LOCKING CONNECTOR WAS RETURNED FOR EVALUATION. NOTE THAT THE TUBING STRAIN RELIEF WAS NOT RETURNED. UPON VISUAL INSPECTION, IT WAS NOTED THAT THE ACTUATOR RING WAS IN THE LOCKED POSITION, THE HOOKS WERE DEPLOYED, AND THE LOCKING CONNECTOR WAS IN THE LOCKED POSITION. BIOLOGICAL DEBRIS WAS OBSERVED AROUND THE ACTUATOR RING AND HOOKS. MICROSCOPIC EVALUATION, SHOWED PUNCTURES ON THE SEPTUM. DIMENSIONAL ANALYSIS WAS PERFORMED AND THE CONNECTION TUBING ARE MEETING SPECIFICATION. IT IS NOTED THAT THE PRODUCT'S INSTRUCTION FOR USE (IFU) PROVIDES SPECIFIC INSTRUCTIONS REGARDING THE CONNECTION METHOD AND ADEQUATE GRASP METHOD FOR THE PORT POSITIONING AND CONNECTIONS. IT WAS THEREFORE TENTATIVELY CONCLUDED THAT FAILURE MAY HAVE RESULTED IN IMPROPER CONNECTION OF THE TUBING TO THE PORT. OUR INVESTIGATIONS CONCLUDED THAT THE DEVICE WAS MANUFACTURED TO SPECIFICATIONS AND MET ALL RELEASE CRITERIA. NO RELEVANT DISCREPANCIES WERE NOTED DURING MANUFACTURING PROCESS. A 100% INSPECTION OCCURS PRIOR TO DEVICE RELEASE.

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION UNAVAILABLE. DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CONSUMER'S SPOUSE THAT POST IMPLANT A REALIZE ADJUSTABLE GASTRIC BAND, THE PORT AND BAND TUBING BECAME DISCONNECTED AND THIS WAS DISCOVERED DURING A BAND FILL UNDER FLUOROSCOPY. A PORT REVISION HAS BEEN SCHEDULED. EXACT DATE UNKNOWN.