FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1954392 · Received December 14, 2010

Report

Report Number
2531779-2010-02844
Event Type
Malfunction
Date Received
December 14, 2010
Date of Event
November 11, 2010
Report Date
November 13, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. EXAMINATION REVEALED NO PHYSICAL DAMAGE OR EVIDENCE OF MOISTURE INGRESS TO THE BATTERY COMPARTMENT; THE BATTERY CAP WAS FOUND TO BE INTACT AND ABLE TO BE SECURED WITHOUT DIFFICULTY. EVAL FOUND THAT PUMP POWERED UP PROPERLY AND THE POWER CIRCUIT DID NOT DRAW EXCESSIVE CURRENT. THE PUMP WAS EXERCISED FOR THREE HOURS AND NO OVERHEATING OR ALARMS WERE DUPLICATED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PUMP AND BATTERY BECAME HOT TO TOUCH. THE PT DENIED THE NEED FOR MEDICAL TREATMENT BECAUSE OF THE TEMP OF THE PUMP. HE STATED THAT THE BATTERY COMPARTMENT IS INTACT, THE BATTERY CAP IS SECURE AND IS TIGHTENED WITH A COIN, AND THERE IS NO CORROSION IN THE BATTERY COMPARTMENT. THE PT STATED THAT HE REPLACED THE BATTERY AND HAS HAD NO FURTHER SIGNS OF OVERHEATING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 52 YR