FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1954349
·
Received December 13, 2010
Report
- Report Number
- 2531779-2010-02847
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Report Date
- October 6, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: NO TRAUMA TO KEYPAD OBSERVED. 'CONTRAST' RESPONDING AND SPRING BACK WHEN PRESSED. DURING A VISUAL INSPECTION, THE 'UP/DOWN/OK' BUTTONS STICKING AND INTERMITTENTLY RESPONSIVE WHEN PRESSED. THERE WAS CONTAMINATION UNDER 'OK/UP/DOWN' BUTTON CONTACTS.
Description of Event or Problem · 1
THE LAY-USER/PT ALLEGED THAT THE BUTTONS ON THE KEYPAD DO NOT RESPOND. THERE WAS PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 |