FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1954325
·
Received December 13, 2010
Report
- Report Number
- 2531779-2010-02848
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Report Date
- October 10, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THERE WAS NO DAMAGE TO KEYPAD OBSERVED DURING VISUAL INSPECTION. THE 'DOWN/OK' BUTTONS INTERMITTENTLY RESPONSIVE WHILE THE 'CONTRAST/UP' BUTTON RESPONSIVE TO USER INPUT. ALL BUTTONS SPRING BACK NORMAL WHEN PRESSED. THERE WAS CONTAMINATION UNDER 'DOWN/OK' BUTTON CONTACTS OBSERVED BUT NO CONTAMINATION WERE FOUND UNDER 'CONTRAST/UP' BUTTON.
Description of Event or Problem · 1
THE LAY-USER/PT ALLEGED THAT THE BUTTONS ON THE KEYPAD DO NOT RESPOND. THERE WAS PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |