FDA Adverse Event Injury Summary report: N

VITALITY

MDR report key: 1954231 · Received January 10, 2011

Report

Report Number
2124215-2010-24270
Event Type
Injury
Date Received
January 10, 2011
Date of Event
December 13, 2010
Report Date
December 13, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. ALL LOW VOLTAGE POWER SUPPLY MEASUREMENTS WERE CURRENTLY WITHIN AN APPROPRIATE RANGE; HOWEVER, FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), WHICH IS PART OF THE SHORTENED REPLACEMENT WINDOW ADVISORY (B)(6) 2007 POPULATION PRODUCT ADVISORY INITIALLY COMMUNICATED ON (B)(6) 2007, WAS EXPLANTED FOR NORMAL BATTERY DEPLETION WITHOUT ALLEGATION OF PREMATURE BATTERY DEPLETION. THERE WAS NO REPORT OF ADVERSE PATIENT EFFECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T135

Patients

Seq Age Sex Outcome Treatment
1 71 YR 0185| T135