VITALITY
Report
- Report Number
- 2124215-2010-24442
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- December 16, 2010
- Report Date
- March 9, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE DEVICE WAS LATER EXPLANTED DUE TO NORMAL BATTERY DEPLETION AND WAS RETURNED FOR ANALYSIS. UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, LABORATORY TECHNICIANS UTILIZED AN ENGINEERING LEVEL LONGEVITY PREDICTION CALCULATION TO ASSESS THE RATE OF BATTERY DEPLETION, GIVEN THE PROGRAMMED PARAMETERS AND OTHER DATA STORED WITHIN THE MEMORY OF THE DEVICE. THE RESULTS OF THIS CALCULATION INDICATED THAT THE ACTUAL RATE OF BATTERY DEPLETION FELL WITHIN AN ACCEPTABLE RANGE. NEXT, A COMPREHENSIVE SERIES OF DIAGNOSTIC TESTS WERE CONDUCTED THAT VERIFIED THE PERFORMANCE OF PACING, SENSING, AND RECORDING FUNCTIONS. HAVING MET THE ENGINEERING LONGEVITY PREDICTION, FUNCTIONALLY PASSING ALL RETURNED PRODUCT TESTING, AND WITH NO FURTHER INFORMATION TO INDICATE A PRODUCT PERFORMANCE ISSUE, WE HAVE CONCLUDED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.
SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS EVENT WILL BE UPDATED AND RESUBMITTED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE DEVICE HAS A MONITORING VOLTAGE OF 2.58 VOLTS AND CHARGE TIME OF 16.9 SECONDS. THIS CHARGE TIME WAS REPORTED AS UNEXPECTED. CURRENTLY THIS DEVICE REMAINS IMPLANTED AND IN SERVICE. NO ADVERSE PATIENT EFFECTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T127 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | 4474| T127| 0186 |