FDA Adverse Event
Malfunction
Summary report: N
ENDOTAK RELIANCE
MDR report key: 1954124
·
Received January 10, 2011
Report
- Report Number
- 2124215-2010-24288
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- December 15, 2010
- Report Date
- December 23, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
AT THIS TIME THE DEVICE AND RIGHT VENTRICULAR LEAD REMAIN IN SERVICE. EFFORTS TO OBTAIN ADDITIONAL INFORMATION WERE MADE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.
Additional Manufacturer Narrative · 1
THE SALES REPRESENTATIVE CONFIRMED THE ISSUE WAS RESOLVED BY CHANGING THE LEAD CONFIGURATION TO DISTAL TO CAN WHICH RESULTED IN SATISFACTORY SHOCK IMPEDANCE MEASUREMENTS. NO FURTHER ISSUES WERE OBSERVED.
Description of Event or Problem · 1
--
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT AN ALERT WAS DETECTED DUE TO HIGH SHOCK IMPEDANCE VALUES. THE REASON FOR THE OUT OF RANGE SHOCK IMPEDANCES WAS NOT PROVIDED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0157 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | 0157| E110| 4469 |