FDA Adverse Event Malfunction Summary report: N

EASYTRAK 2

MDR report key: 1953990 · Received January 10, 2011

Report

Report Number
2124215-2010-24145
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
December 14, 2010
Report Date
December 14, 2010
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED LOW SENSING MEASUREMENTS AND LOSS OF CAPTURE WITH AN IMPEDANCE MEASUREMENT GREATER THAN 2000 OHMS. THE PHYSICIAN OPTED TO ELECTRICALLY ABANDON THE LV LEAD. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4542

Patients

Seq Age Sex Outcome Treatment
1 84 YR (B)(4)| (B)(4)| (B)(4)| (B)(4)| (B)(4)| (B)(4)| (B)(4)