FDA Adverse Event Malfunction Summary report: N

ALTRUA

MDR report key: 1953964 · Received January 10, 2011

Report

Report Number
2124215-2010-24084
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
October 12, 2010
Report Date
June 19, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS ELECTIVELY EXPLANTED OVER THREE YEARS AFTER THE INITIAL OBSERVATIONS WERE REPORTED. THE DEVICE WAS RETURNED AND IS BEING EVALUATED IN OUR POST MARKET QUALITY ASSURANCE LABORATORY. THIS PRODUCT ISSUE WILL BE UPDATED WHEN ANALYSIS IS COMPLETE.

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE RIGHT VENTRICULAR PACING, AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. LABORATORY ANALYSIS DID NOT IDENTIFY ANY DEVICE CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

THE DEVICE REMAINS ACTIVELY IMPLANTED AND NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENT WAS UNDERGOING AN UNSPECIFIED MEDICAL PROCEDURE AND THERE WAS INTERACTION BETWEEN THE RESPIRATORY RATE MONITOR AND THE MINUTE VENTILATION (MV) SENSORS ON THE PACEMAKER. PACING AT THE MAXIMUM SENSOR RATE (MSR) WAS OBSERVED. THE MV FEATURE WAS PROGRAMMED OFF UNTIL THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE AFFECTS FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALTRUA IMPLANTABLE PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND S603

Patients

Seq Age Sex Outcome Treatment
1 67 YR 4469| 4470| S603