FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 1953951 · Received January 10, 2011

Report

Report Number
2124215-2010-23926
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
December 13, 2010
Report Date
December 21, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR (B)(4) LABORATORY, DEVICE INTERROGATION REVEALED THAT THE DEVICE BATTERY STATUS WAS ELECTIVE REPLACEMENT INDICATOR (ERI), WITH A BATTERY VOLTAGE OF 2.34 VOLTS. A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO A COMPROMISED LOW-VOLTAGE CAPACITORS, WHICH RESULTED IN A HIGH CURRENT CONDITION.

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED UPON COMPLETION OF THE ANALYSIS.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE, WHICH WAS INCLUDED IN THE SHORTENED REPLACEMENT WINDOW ADVISORY (B)(6) 2007 POPULATION PRODUCT ADVISORY WAS INITIALLY COMMUNICATED ON (B)(6) 2007, WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. TO DATE, NO ADVERSE PATIENT EFFECTS WERE REPORTED WITH THIS CLINICAL OBSERVATION. THE DEVICE WAS RETURNED FOR RELIABILITY ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H170

Patients

Seq Age Sex Outcome Treatment
1 73 YR H170| H227| 4543| 4086| 0157