FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 1953945 · Received January 10, 2011

Report

Report Number
2124215-2010-24043
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
December 15, 2010
Report Date
December 21, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Removal / Correction Number
Z-1620-09 TO Z-1621-09
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, ICD MEMORY WAS REVIEWED. WE CONFIRMED THAT THE DEVICE DECLARED ERI AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT OF 18.9 SECONDS. TO DETERMINE IF THE DEVICE'S RATE OF BATTERY DEPLETION WAS NORMAL, OUR TECHNICIANS COMPARED THE OBSERVED RATE OF BATTERY USAGE TO THE EXPECTED RATE OF BATTERY USAGE. THE RESULTS SHOWED THAT THE MONITORING VOLTAGE WAS NORMAL BASED ON THERAPY USE TO DATE AND PROGRAMMED SETTINGS. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY CHARGE TIMES IN EXCESS OF THE 18.9 SECOND EXTENDED ERI CHARGE TIME LIMIT. LABORATORY TECHNICIANS AND ENGINEERS CONCLUDED THAT THIS DEVICE DID NOT EXPERIENCE PREMATURE BATTERY DEPLETION. RATHER, AN ERI CHARGE TIME REPLACEMENT INDICATOR WAS DECLARED DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Additional Manufacturer Narrative · 1

UPON COMPLETION OF THE ANALYSIS, THIS EVENT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS REMOVED AND RETURNED. THERE WAS NO CLINICAL ALLEGATION AGAINST THIS DEVICE. INITIAL ANALYSIS REVEALED THIS DEVICE REACHED ELECTIVE REPLACEMENT INDICATORS (ERI) WHILE IMPLANTED AND DID NOT MEET EXPECTED LONGEVITY CALCULATIONS. FURTHER ANALYSIS WILL BE PERFORMED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T167

Patients

Seq Age Sex Outcome Treatment
1 75 YR 4470| 0158| T167| MISMATCH