FDA Adverse Event Injury Summary report: N

VENTAK PRIZM

MDR report key: 1953942 · Received January 10, 2011

Report

Report Number
2124215-2010-24057
Event Type
Injury
Date Received
January 10, 2011
Date of Event
December 14, 2010
Report Date
April 20, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS ICD REMAINS IMPLANTED, SO THEREFORE WILL NOT BE RETURNED TO BOSTON SCIENTIFIC FOR LABORATORY ANALYSIS. AT THIS TIME THERE IS NO ADDITIONAL INFORMATION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

SUBSEQUENTLY, THIS ICD WAS RECEIVED AT OUR POST MARKET QUALITY ASSURANCE LABORATORY. UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, LABORATORY TECHNICIANS USED AN ENGINEERING LEVEL LONGEVITY PREDICTION CALCULATION TO ASSESS THE RATE OF BATTERY DEPLETION, GIVEN THE PROGRAMMED PARAMETERS AND OTHER DATA STORED WITHIN THE MEMORY OF THE DEVICE. THE RESULTS OF THIS CALCULATION INDICATED THAT THE ACTUAL RATE OF BATTERY DEPLETION FELL WITHIN AN ACCEPTABLE RANGE. NEXT, A COMPREHENSIVE SERIES OF DIAGNOSTIC TESTS WERE CONDUCTED THAT VERIFIED THE PERFORMANCE OF PACING, SENSING, AND RECORDING FUNCTIONS. HAVING MET THE ENGINEERING LONGEVITY PREDICTION, FUNCTIONALLY PASSING ALL RETURNED PRODUCT TESTING, AND WITH NO FURTHER INFORMATION TO INDICATE A PRODUCT PERFORMANCE ISSUE, WE HAVE CONCLUDED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). IT WAS SUSPECTED THE BATTERY DEPLETED EARLIER THAN EXPECTED. THE DECISION WAS MADE TO EXPLANT AND REPLACE THIS DEVICE WITHIN THE NEAR FUTURE. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VENTAK PRIZM IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND 1853

Patients

Seq Age Sex Outcome Treatment
1