VENTAK PRIZM
Report
- Report Number
- 2124215-2010-24094
- Event Type
- Injury
- Date Received
- January 10, 2011
- Date of Event
- December 15, 2010
- Report Date
- January 14, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, LABORATORY TECHNICIANS UTILIZED AN ENGINEERING LEVEL LONGEVITY PREDICTION CALCULATION TO ASSESS THE RATE OF BATTERY DEPLETION, GIVEN THE PROGRAMMED PARAMETERS AND OTHER DATA STORED WITHIN THE MEMORY OF THE DEVICE. THE RESULTS OF THIS CALCULATION INDICATED THAT THE ACTUAL RATE OF BATTERY DEPLETION FELL WITHIN AN ACCEPTABLE RANGE. NEXT, A COMPREHENSIVE SERIES OF DIAGNOSTIC TESTS WERE CONDUCTED THAT VERIFIED THE PERFORMANCE OF PACING, SENSING, AND RECORDING FUNCTIONS. HAVING MET THE ENGINEERING LONGEVITY PREDICTION, FUNCTIONALLY PASSING ALL RETURNED PRODUCT TESTING, AND WITH NO FURTHER INFORMATION TO INDICATE A PRODUCT PERFORMANCE ISSUE, WE HAVE CONCLUDED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE WILL BE RETURNED FOR ANALYSIS. UPON COMPLETION OF THE ANALYSIS, THIS EVENT WILL BE UPDATED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT A REPLACEMENT PROCEDURE WAS PERFORMED. THIS DEVICE WAS REMOVED AND REPLACED. THERE WAS CONCERN THAT THE BATTERY DEPLETED MORE RAPIDLY THAN EXPECTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
- -
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VENTAK PRIZM | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | 1853 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |