FDA Adverse Event
Malfunction
Summary report: N
HEARTSTART MRX
MDR report key: 1953903
·
Received December 15, 2010
Report
- Report Number
- 1218950-2010-02605
- Event Type
- Malfunction
- Date Received
- December 15, 2010
- Report Date
- November 17, 2010
- Manufacturer
- PHILIPS HEALTHCARE
- Product Code
- MKJ
- PMA / PMN Number
- K031187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4): THE CUSTOMER REPORTED THAT THE DEVICE INTERMITTENTLY DISPLAYED A RED X. THE DEVICE WAS EVALUATED AT PHILIPS. THE SYMPTOM WAS FURTHER CLARIFIED TO BE AN INTERMITTENT HANG ON POWER UP. THE DEVICE WAS REPAIRED BY REPLACEMENT OF THE PROCESSOR PCA. THE DEVICE PASSED ALL TESTING AND WAS RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE DEVICE INTERMITTENTLY DISPLAYED A RED X.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSTART MRX | MKJ | PHILIPS HEALTHCARE | M3535A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |