FDA Adverse Event Malfunction Summary report: N

HEARTSTART MRX

MDR report key: 1953903 · Received December 15, 2010

Report

Report Number
1218950-2010-02605
Event Type
Malfunction
Date Received
December 15, 2010
Report Date
November 17, 2010
Manufacturer
PHILIPS HEALTHCARE
Product Code
MKJ
PMA / PMN Number
K031187
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE CUSTOMER REPORTED THAT THE DEVICE INTERMITTENTLY DISPLAYED A RED X. THE DEVICE WAS EVALUATED AT PHILIPS. THE SYMPTOM WAS FURTHER CLARIFIED TO BE AN INTERMITTENT HANG ON POWER UP. THE DEVICE WAS REPAIRED BY REPLACEMENT OF THE PROCESSOR PCA. THE DEVICE PASSED ALL TESTING AND WAS RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE DEVICE INTERMITTENTLY DISPLAYED A RED X.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTART MRX MKJ PHILIPS HEALTHCARE M3535A

Patients

Seq Age Sex Outcome Treatment
1