BD PEGASUS GN 18GAX1.16IN PRN-CAP Y NON-PVC
Report
- Report Number
- 3002601200-2024-00254
- Event Type
- Malfunction
- Date Received
- June 14, 2024
- Date of Event
- May 23, 2024
- Report Date
- July 2, 2024
- Manufacturer
- BD SUZHOU (MDS)
- Product Code
- FOZ
- UDI-DI
- 00382903837564
- PMA / PMN Number
- UNK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
1.DHR REVIEW : 1 THE BATCH NUMBER OF THE COMPLAINED PRODUCT IS 3234106, IS 18G AND PRODUCT CODE IS 383756, PRODUCED ON 2023/09, WITH A TOTAL OF 19000 PIECES IN THIS BATCH 2 INSPECTION PROCESS INSPECTION AND DELIVERY INSPECTION REPORT, THE TEST RESULTS MEET THE PRODUCT STANDARDS, NO ABNORMALITY. 3 CHECK THE PRODUCTION RECORDS FOR THIS BATCH OF PRODUCTS THAT THERE ARE NO NONCONFORMITIES, DEVIATIONS OR REWORK ACTIVITIES IN THE PROCESS OF THIS BATCH OF PRODUCTS. 2.THE CUSTOMER DID NOT RETURN SAMPLES OR PHOTOS; THE DEFECT STATUS CANNOT BE CONFIRMED. 3. TAKE THE RETAINED SAMPLE OF THIS BATCH FOR 45PSI SYSTEM LEAKAGE TEST, AND NO ABNORMALITY WAS FOUND. 4. THE HISTORY OF CUSTOMER COMPLAINTS FOR THE SAME BATCH OF PRODUCTS HAS BEEN REVIEWED, THIS COMPLAINT IS THE SECOND TIME, AND THE RELEVANT COMPLAINT WAS (B)(4). IN SUMMARY, SINCE THE CUSTOMER DOES NOT HAVE SAMPLES OR PHOTOS RETURNED, THE SPECIFIC DEFECT STATUS CANNOT BE CONFIRMED. THEREFORE, THE ROOT CAUSE OF THE COMPLAINED DEFECT CANNOT BE DETERMINED. THE FACTORY WILL CONTINUE TO PAY ATTENTION TO AND MONITOR THE DEFECT COMPLAINT TREND. IN SUMMARY, THE ROOT CAUSE OF THE COMPLAINT DEFECT CANNOT BE CONFIRMED, AND THE FACTORY WILL CONTINUE TO PAY ATTENTION TO AND MONITOR THE COMPLAINT TREND OF THE DEFECT.
IT WAS REPORTED THAT BD PEGASUS GN 18GAX1.16IN PRN-CAP Y NON-PVC LEAKED. AFTER PERFORMING THE PUNCTURE, THE NURSE FOUND THAT THE LIQUID WAS COMING OUT OF THE ISOLATION PLUG.
NO ADDITIONAL INFORMATION REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 190458 | BD PEGASUS GN 18GAX1.16IN PRN-CAP Y NON-PVC | INTRAVASCULAR CATHETER | FOZ | BD SUZHOU (MDS) | 3234106 | 00382903837564 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |