FDA Adverse Event Malfunction Summary report: N

TELIGEN

MDR report key: 1953687 · Received January 10, 2011

Report

Report Number
2124215-2010-23889
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
October 21, 2010
Report Date
February 15, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WERE ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Additional Manufacturer Narrative · 1

THE DEVICE WAS LATER RETURNED. THE REASON FOR EXPLANT WAS NOT DISCLOSED AS RECORDS INDICATED THE DEVICE REMAINED IMPLANTED.

Additional Manufacturer Narrative · 1

INFORMATION SUGGESTS THE LEAD AND DEVICE REMAIN IN-SERVICE. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS MADE AVAILABLE THAT THIS DEVICE WAS EXPLANTED DUE TO AN INFECTION, THIS INFORMATION IS UNRELATED TO THIS REPORT AND IS COVERED IN REPORT NUMBER 2124215-2011-04019.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DETECTED OUT OF RANGE SHOCK IMPEDANCES ON THIS DEFIBRILLATION LEAD. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND E110

Patients

Seq Age Sex Outcome Treatment
1 54 YR 4136| E110| 0180