ALTRUA
Report
- Report Number
- 2124215-2010-23898
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- December 13, 2010
- Report Date
- February 4, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- PMA / PMN Number
- D970003
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER EXHIBITED RIGHT VENTRICULAR LOSS OF CAPTURE DURING AN IMPLANT PROCEDURE. ANOTHER PACEMAKER WITH LOSS OF CAPTURE BEEN ATTEMPTED DURING THE SAME PROCEDURE. THE RELATED LEAD HAD GOOD THRESHOLD MEASUREMENTS AS MEASURED THROUGH THE PACING SYSTEM ANALYZER (PSA) AND MULTIPLE TROUBLESHOOTING STEPS WERE TAKEN. HOWEVER, NO CAPTURE WAS NOTED. THE LEAD WAS THEN EXPLANTED AND THE PROCEDURE ABANDONED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALTRUA | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | S601 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |