FDA Adverse Event Malfunction Summary report: N

COGNIS

MDR report key: 1953619 · Received January 10, 2011

Report

Report Number
2124215-2010-24561
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
December 9, 2010
Report Date
October 15, 2012
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
PMA / PMN Number
P010012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED AN IMPEDANCE MEASUREMENT GREATER THAN 125 OHMS THAT WAS ABLE TO BE REPRODUCED WHEN THE PATIENT RAISED THEIR ARMS. THE PHYSICIAN IS STILL CONTEMPLATING IF HE WILL PERFORM A LEAD REVISION PROCEDURE. IT WAS REPORTED THAT THE HEALTH CARE PROFESSIONAL (HCP) ELECTED TO TURN OFF DAILY MEASUREMENTS FOR THE HIGH SHOCK IMPEDANCE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT A COMPUTERIZED TOMOGRAPHY (CT) SCAN REVEALED THAT THE LEAD HAD LIKELY PERFORATED THE RIGHT VENTRICLE. A REVISION PROCEDURE WAS PERFORMED WHERE A PORTION OF THE EXISTING RV LEAD WAS SURGICALLY ABANDONED DUE TO THE HIGH OUT OF RANGE SHOCK IMPEDANCE MEASUREMENTS AND SUSPECTED PERFORATION. SINCE RV THRESHOLD AND SENSING MEASUREMENTS HAVE BEEN STABLE AND THE LEAD WAS IMPLANTED APPROXIMATELY EIGHT YEARS EARLIER, IT IS UNKNOWN WHEN THE PERFORATION OCCURRED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED AN IMPEDANCE MEASUREMENT GREATER THAN 125 OHMS THAT WAS ABLE TO BE REPRODUCED WHEN THE PATIENT RAISED THEIR ARMS. THE PHYSICIAN IS STILL CONTEMPLATING IF HE WILL PERFORM A LEAD REVISION PROCEDURE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COGNIS IMPLANTABLE CHF PULSE GENERATOR NIK GUIDANT CRM CLONMEL IRELAND N119

Patients

Seq Age Sex Outcome Treatment
1 57 YR H215| 5076| 0185| H175| H170| 4046| 4513| N119| 4470