FDA Adverse Event Malfunction Summary report: N

VITALITY

MDR report key: 1953554 · Received January 10, 2011

Report

Report Number
2124215-2010-23688
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
December 8, 2010
Report Date
December 8, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILATOR (ICD) WAS RETURNED FOR DISPOSAL. THERE WERE NO ALLEGATIONS AGAINST THE FUNCITIONALITY OR LONGEVITY OF THIS DEVICE. NO ADVERSE PATIENT EFFECTS REPORTED. NO LONGER IN SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T127

Patients

Seq Age Sex Outcome Treatment
1 74 YR Other 0184| 4469| T127