FDA Adverse Event Injury Summary report: N

ANEURX AAADVANTAGE STENT GRAFT SYSTEM-HYDRO

MDR report key: 1953484 · Received January 5, 2011

Report

Report Number
2953200-2011-00014
Event Type
Injury
Date Received
January 5, 2011
Date of Event
December 6, 2010
Report Date
December 6, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION, RESULTS: (ENDOLEAK). RESULTS/CONCLUSIONS: (CALCIFIED AORTIC NECK).

Description of Event or Problem · 1

AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. THE ANEURYSM WAS 5.5 CM IN DIAMETER; THE AORTIC NECK 19 MM IN DIAMETER FOR THE 12 MM IN LENGTH OF THE AORTIC NECK. THE VESSEL MORPHOLOGY WAS REPORTED AS SEVERE CALCIFICATION AND MODERATE TORTUOSITY AND THERE WAS 20 DEGREE ANGULATION IN THE AORTIC NECK. THE STENT GRAFTS WERE IMPLANTED UPON FINAL ANGIOGRAM THERE WAS A PROXIMAL TYPE I ENDOLEAK PRESENT, DUE TO CALCIFICATION IN THE AORTIC NECK. THE PHYSICIAN ELECTED TO IMPLANT AN ANEURX AORTIC CUFF THE ENDOLEAK DECREASED; HOWEVER, DID NOT COMPLETELY RESOLVE. THE PHYSICIAN ELECTED TO MONITOR THE PATIENT. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX AAADVANTAGE STENT GRAFT SYSTEM-HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00526488

Patients

Seq Age Sex Outcome Treatment
1 89 YR Required Intervention