ANEURX AAADVANTAGE STENT GRAFT SYSTEM-HYDRO
Report
- Report Number
- 2953200-2011-00014
- Event Type
- Injury
- Date Received
- January 5, 2011
- Date of Event
- December 6, 2010
- Report Date
- December 6, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION, RESULTS: (ENDOLEAK). RESULTS/CONCLUSIONS: (CALCIFIED AORTIC NECK).
AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. THE ANEURYSM WAS 5.5 CM IN DIAMETER; THE AORTIC NECK 19 MM IN DIAMETER FOR THE 12 MM IN LENGTH OF THE AORTIC NECK. THE VESSEL MORPHOLOGY WAS REPORTED AS SEVERE CALCIFICATION AND MODERATE TORTUOSITY AND THERE WAS 20 DEGREE ANGULATION IN THE AORTIC NECK. THE STENT GRAFTS WERE IMPLANTED UPON FINAL ANGIOGRAM THERE WAS A PROXIMAL TYPE I ENDOLEAK PRESENT, DUE TO CALCIFICATION IN THE AORTIC NECK. THE PHYSICIAN ELECTED TO IMPLANT AN ANEURX AORTIC CUFF THE ENDOLEAK DECREASED; HOWEVER, DID NOT COMPLETELY RESOLVE. THE PHYSICIAN ELECTED TO MONITOR THE PATIENT. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANEURX AAADVANTAGE STENT GRAFT SYSTEM-HYDRO | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00526488 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 89 YR | Required Intervention |