FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 1953428 · Received January 10, 2011

Report

Report Number
2050012-2011-00039
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
December 16, 2010
Report Date
December 16, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K060256
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER TECHNICAL SUPPORT (CTS) ASSISTED THE CUSTOMER WITH TROUBLESHOOTING AND THE CUSTOMER VERIFIED THAT THE SYRINGE WAS TIGHT. SERVICE REPLACED THE WASH SYRINGE VALVE AND PRIMED THE SYSTEM. THE ISSUE HAS BEEN RESOLVED. (B)(6).

Additional Manufacturer Narrative · 1

THE CUSTOMER TECHNICAL SUPPORT (CTS) ASSISTED THE CUSTOMER WITH TROUBLESHOOTING AND THE CUSTOMER VERIFIED THAT THE SYRINGE WAS TIGHT. SERVICE REPLACED THE WASH SYRINGE VALVE AND PRIMED THE SYSTEM. THE ISSUE HAS BEEN RESOLVED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO CLOSE TUBE ACCESSION (CTA) LEAKING WASH BUFFER. NO INJURY AND/OR PERSONNEL EXPOSURE TO CHEMICAL OR BIO-HAZARDOUS MATERIAL WAS REPORTED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO CLOSE TUBE ACCESSION (CTA) LEAKING WASH BUFFER. NO INJURY AND/OR PERSONNEL EXPOSURE TO CHEMICAL OR BIO-HAZARDOUS MATERIAL WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 600I N/A

Patients

Seq Age Sex Outcome Treatment
1