FDA Adverse Event Malfunction Summary report: N

SYNCHRON® LX® I 725 CLINICAL SYSTEM

MDR report key: 1953418 · Received January 10, 2011

Report

Report Number
2050012-2011-00072
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
December 13, 2010
Report Date
December 13, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K023049
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

A BCI FIELD SERVICE ENGINEER (FSE) FOUND WASH CONCENTRATE SOLUTION SPILLED IN THE HYDROPNEUMATIC DRAWER. FSE REPLACED THE WASH CONCENTRATE FLOAT SWITCH AND VALVES V12 AND V14.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO HYDROPNEUMATIC COMPARTMENT LEAK. NO INJURY WAS REPORTED FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHRON® LX® I 725 CLINICAL SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. LXI 725 N/A

Patients

Seq Age Sex Outcome Treatment
1