FDA Adverse Event
Malfunction
Summary report: N
SYNCHRON® LX® I 725 CLINICAL SYSTEM
MDR report key: 1953418
·
Received January 10, 2011
Report
- Report Number
- 2050012-2011-00072
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- December 13, 2010
- Report Date
- December 13, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K023049
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
Additional Manufacturer Narrative · 1
A BCI FIELD SERVICE ENGINEER (FSE) FOUND WASH CONCENTRATE SOLUTION SPILLED IN THE HYDROPNEUMATIC DRAWER. FSE REPLACED THE WASH CONCENTRATE FLOAT SWITCH AND VALVES V12 AND V14.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO HYDROPNEUMATIC COMPARTMENT LEAK. NO INJURY WAS REPORTED FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON® LX® I 725 CLINICAL SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | LXI 725 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |