FDA Adverse Event Malfunction Summary report: N

ARCHITECT STAT TROPONIN-I

MDR report key: 1953374 · Received January 8, 2011

Report

Report Number
1415939-2011-00006
Event Type
Malfunction
Date Received
January 8, 2011
Date of Event
November 27, 2010
Report Date
December 13, 2010
Manufacturer
ABBOTT LABORATORIES
Product Code
MMI
PMA / PMN Number
K041192
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A STUDY WAS CONDUCTED USING NINE LEVELS OF HUMAN SERUM PANELS PREPARED TO CONCENTRATIONS RANGING FROM 0.01 NG/ML TO 0.20 NG/ML. PANELS WERE TESTED IN REPLICATES OF 2 TWICE EACH DAY FOR A PERIOD OF 5 DAYS ACROSS 2 INSTRUMENTS. USING AN IN-HOUSE FILE KIT OF REAGENT LOT 39483UN10, A TOTAL OF 20 REPLICATES PER PANEL PER INSTRUMENT WERE TESTED. THE %CV'S OF EACH PANEL WERE CALCULATED AND PLOTTED AGAINST THE MEAN CONCENTRATIONS OF EACH PANEL. A CURVE WAS THEN FITTED THROUGH THE DATA POINTS AND THE 10% CV VALUE WAS ESTIMATED AS THE CONCENTRATION CORRESPONDING TO THE 10% CV LEVEL OF THE FITTED CURVE. THE TEST RESULTS MET THE ASSAY'S PACKAGE INSERT CLAIM, THAT THE ARCHITECT STAT TROPONIN-I CONCENTRATION AT 10% CV IS LESS THAN OR EQUAL TO 0.10 NG/ML. IN ADDITION TO TESTING, COMPLAINTS WERE REVIEWED FROM (B)(4) 2010 THROUGH (B)(4) 2011 TO DETERMINE IF OTHERS HAVE EXPERIENCED SIMILAR ISSUES THAT MIGHT WARRANT FURTHER INVESTIGATION. THIS REVIEW DID NOT IDENTIFY AN ADVERSE TREND RELATED TO THE CUSTOMER'S OBSERVATION. BASED ON THE RESULTS OF THIS INVESTIGATION, A PRODUCT DEFICIENCY WAS NOT IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4). AN INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER HAS OBSERVED IMPRECISION ISSUES WITH THE ARCHITECT STAT TROPONIN-I ASSAY ACROSS TWO I2000SR ANALYZERS WHERE THE ORIGINAL RESULT DOES NOT REPEAT. THE CUSTOMER HAS AN APS TRACK SYSTEM IN THEIR LABORATORY WITH TWO I2000SR ANALYZERS ATTACHED. AN ALGORITHM HAS BEEN SET UP IN THEIR INSTRUMENT MANAGER SOFTWARE AND RESULTS BETWEEN 0.03 AND 0.3 NG/ML ARE AUTOMATICALLY REPEATED. THE CUSTOMER PROVIDED PATIENT DATA GENERATED BETWEEN (B)(6) 2010. FOURTEEN PATIENT SAMPLES HAD DISCREPANT VALUES THAT CROSSED THE CUSTOMER'S CUT-OFF OF 0.03 NG/ML. THIS REPORT IS FOR PATIENT #5 OF 14 SIMILAR IMPRECISION REPORTS. THE INITIAL RESULT WAS 0.038 NG/ML AND UPON REPEAT, YIELDED RESULTS OF 0.018 AND 0.034 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT STAT TROPONIN-I FOR THE DETERMINATION OF CARDIAC TROPONIN-I IN HUMAN SERUM OR PLASMA MMI ABBOTT LABORATORIES 39483UN10

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT I2000SR LIST # 3M74-01 SERIAL # (B)(4)| ARCHITECT I2000SR LIST # 3M74-01 SERIAL # (B)(4)