FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM
MDR report key: 1953342
·
Received January 8, 2011
Report
- Report Number
- 2050012-2011-00017
- Event Type
- Malfunction
- Date Received
- January 8, 2011
- Date of Event
- December 9, 2010
- Report Date
- December 9, 2010
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BCI FIELD SERVICE ENGINEER (FSE) CLEANED THE SPILL WHICH WAS NO FOAM SOLUTION. THE FSE REPLACED THE TUBING CONNECTED FROM THE SOLENOIDS TO THE NO FOAM BOTTLE.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER, INC. (BCI) IN REGARDS TO A FLUID LEAK AROUND THE WASH CONCENTRATE BOTTLE AREA OF UNICEL DXC 600 PRO SYNCHRON CLINICAL SYSTEM. NO INJURY WAS REPORTED AND THERE WAS NO EFFECT TO PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM | CLINICAL CHEMSTRY ANALYZER | JJE | BECKMAN COULTER, INC. | DXC 600 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |