FDA Adverse Event Malfunction Summary report: N

SYNCHRON UNICEL DXC 800 PRO CLINICAL SYSTEMS

MDR report key: 1953325 · Received January 7, 2011

Report

Report Number
2050012-2011-00061
Event Type
Malfunction
Date Received
January 7, 2011
Date of Event
September 20, 2010
Report Date
September 20, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JGS
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ION-SELECTIVE ELECTRODE (ISE) SYSTEM IS CALIBRATED EVERY 24 HOURS. ISE QC SAMPLES ARE RUN EVERY 8 HOURS. QC RESULTS WERE WITHIN THE LAB'S ESTABLISHED RANGES. UNDER THE DIRECTION OF THE HOTLINE, THE CUSTOMER CONFIRMED SEEING BUBBLE IN THE RATIO PUMP ON THE FIRST PRIME CYCLE. A FIELD SERVICE ENGINEER (FSE) COMPLETED ISE PREVENTATIVE MAINTENANCE (PM) WHICH INCLUDED: BLEACHING THE ISE SYSTEM REPLACING ALL ISE TUBING, CARBON BRIDGE, AND ELECTROLYTE INJECTION CUP (EIC) SEALS REBUILDING THE RATIO PUMP. NO FURTHER CALLS HAVE BEEN MADE TO THE BCI HOTLINE IN REGARDS TO THE ISE PROBLEMS. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS WITHIN THE DATE RANGE (B)(6) 2010 - (B)(6) 2010 FOR ADDITIONAL REPORTABLE EVENTS/OCCURRENCES. THIS REPORTABLE EVENT IS RELATED TO PREVIOUSLY SUBMITTED MDR 2050012-2010-00974.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING ERRONEOUS HIGH SODIUM (NA), POTASSIUM (K), CHLORIDE (CL), CARBON DIOXIDE (CO2), AND CALCIUM (CA) RESULTS GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THE SAMPLE WAS REPEATED AND LOWER ISE RESULTS WERE OBTAINED. ALL RESULTS OF ELECTROLYTES IN THIS LAB WERE GENERATED FROM SYSTEM ID (B)(4) AND (B)(4). SYSTEM ID (B)(4) IS CAPTURED IN MDR 2050012-2010-00974. PATIENT TREATMENT WAS NOT IMPACTED BY THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHRON UNICEL DXC 800 PRO CLINICAL SYSTEMS CLINICAL CHEMISTRY ANALYZER JGS BECKMAN COULTER INC. DXC 800 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1