FDA Adverse Event Malfunction Summary report: N

HOMECHOICE CYCLER-REFURBISHED

MDR report key: 1953295 · Received January 7, 2011

Report

Report Number
1423500-2011-00300
Event Type
Malfunction
Date Received
January 7, 2011
Date of Event
November 2, 2010
Report Date
December 15, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).REVIEW OF THE DEVICE LOGS CONFIRMED THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. EXTERNAL & INTERNAL VISUAL INSPECTIONS REVEALED NO PROBLEMS. THE CAUSE OF THE IIPV WAS DETERMINED TO BE INSUFFICIENT DRAIN DUE TO ONE OR MORE CYCLE ADVANCING TO THE NEXT FILL WHEN SLOW/NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. A SERVICE HISTORY REVIEW (SHR) REVEALED THAT NO ISSUES THAT MAY HAVE CONTRIBUTED TO THE IIPV. THE DEVICE WAS DETERMINED TO MEET THE SPECIFICATION REQUIREMENTS RELATIVE TO THE IIPV IDENTIFIED VIA DEVICE LOG REVIEW. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). TWO OF 3 EMDRS.THE DEVICE HAS BEEN RECEIVED AT BAXTER FOR EVALUATION. HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED ON (B)(6) 2010 DURING DRAIN CYCLE 2. THE PATIENT'S LARGEST PRESCRIBED FILL VOLUME (LPFV) WAS 490 ML AND THE DRAIN VOLUME WAS 1042 ML, WHICH MET IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE CYCLER-REFURBISHED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 27 YR