HOMECHOICE AUTOMATED PD SET W/CASSETTE - 4 PRONG
Report
- Report Number
- 1423500-2011-00284
- Event Type
- Malfunction
- Date Received
- January 7, 2011
- Date of Event
- November 20, 2010
- Report Date
- December 14, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER, THEREFORE, THE DEVICE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WAS NOT BE PERFORMED. NOT ENOUGH DATA IS AVAILABLE WITHIN THE COMPLAINT TO IDENTIFY ROOT CAUSE; THEREFORE, THE CAUSE OF THE COMPLAINT WAS NOT DETERMINED. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
A SYSTEM ERROR (SE) 2240 ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE (HC) DEVICE; OCCURRENCE DATE (B)(6) 2010 DURING INITIAL DRAIN. A SE 2240 (AIR IN SET) IS A NON-RECOVERABLE ALARM THAT OCCURS WHEN THE DEVICE HAS DETECTED AIR BEING DRAWN CONTINUOUSLY INTO THE DISPOSABLE. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET W/CASSETTE - 4 PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | HOMECHOICE CYCLER |