COBAS 8000 C 702 MODULE
Report
- Report Number
- 1823260-2024-01765
- Event Type
- Malfunction
- Date Received
- June 13, 2024
- Date of Event
- May 14, 2024
- Report Date
- August 12, 2024
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K100853
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE FIELD SERVICE ENGINEER (FSE) FOUND THAT THE DETERGENT CONCENTRATION WAS HIGH. THE FSE REPLACED TWO SOLENOID VALVES AND A CHECK VALVE AND THE DETERGENT CONCENTRATION BECAME NORMAL. THE SERVICE MAINTENANCE ACTIONS PERFORMED BY THE FSE RESOLVED THE ISSUE.
THE REAGENT LOT NUMBERS AND EXPIRATION DATES WERE NOT PROVIDED. THE FIELD SERVICE ENGINEER (FSE) FOUND A LEAKING CELL RINSE TUBE AND REPLACED IT. THE INVESTIGATION IS ONGOING.
THERE WAS AN ALLEGATION OF QUESTIONABLE CA2 CALCIUM GEN.2 AND CARB3 CARBAMAZEPINE RESULTS FOR 15 PATIENT SAMPLES ON A COBAS 8000 C702 MODULE. FOR SAMPLE (B)(6), THE INITIAL CA2 RESULT WAS 1.000 MMOL/L. THE REPEAT RESULT WAS 2.020 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CA2 RESULT WAS 0.860 MMOL/L. THE REPEAT RESULT WAS 2.020 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CA2 RESULT WAS 1.910 MMOL/L. THE REPEAT RESULT WAS 2.560 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CA2 RESULT WAS 1.450 MMOL/L. THE REPEAT RESULT WAS 2.510 MMOL/L. FOR SAMPLE 5, THE INITIAL CA2 RESULT WAS 1.580 MMOL/L. THE REPEAT RESULT WAS 2.540 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CA2 RESULT WAS 1.410 MMOL/L. THE REPEAT RESULT WAS 2.760 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CARB3 RESULT WAS 3.400 MMOL/L. THE REPEAT RESULT WAS 4.700 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CARB3 RESULT WAS 2.800 MMOL/L. THE REPEAT RESULT WAS 3.600 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CARB3 RESULT WAS 3.500 MMOL/L. THE REPEAT RESULT WAS 5.700 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CARB3 RESULT WAS 1.800 MMOL/L. THE REPEAT RESULT WAS 2.200 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CARB3 RESULT WAS 3.900 MMOL/L. THE REPEAT RESULT WAS 4.900 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CARB3 RESULT WAS 2.100 MMOL/L. THE REPEAT RESULT WAS 4.300 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CARB3 RESULT WAS 3.400 MMOL/L. THE REPEAT RESULT WAS 5.800 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CARB3 RESULT WAS 2.700 MMOL/L. THE REPEAT RESULT WAS 3.600 MMOL/L. FOR SAMPLE (B)(6), THE INITIAL CARB3 RESULT WAS 5.200 MMOL/L. THE REPEAT RESULT WAS 8.800 MMOL/L.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 766711 | COBAS 8000 C 702 MODULE | CLINICAL CHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |