FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 1953159 · Received January 7, 2011

Report

Report Number
2024168-2011-00116
Event Type
Injury
Date Received
January 7, 2011
Date of Event
November 18, 2010
Report Date
December 15, 2010
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION OF THE RETURNED DEVICE FOUND THE PLUNGER, CUFFS, LINK, NEEDLE TIP AND MONOFILAMENT WERE NOT RETURNED. THERE WAS NO NEEDLE STRIKE MARK ON THE FOOT. THERE WAS NO ABNORMAL OBSERVATION FOUND ON THE RETURNED DEVICE THAT COULD HAVE CONTRIBUTED TO THE REPORTED EVENT. DURING THE INVESTIGATION, A PROXY PLUNGER WAS INSERTED AND THE NEEDLE TRAJECTORY WAS ACCEPTABLE. ALSO, THE NEEDLE TRAJECTORY OF EVERY DEVICE IS CHECKED TWICE DURING MANUFACTURING. THE MOST PROBABLE ROOT CAUSE FOR THE REPORTED CUFF MISS IS NEEDLE DEFLECTION DURING PLUNGER DEPLOYMENT DUE TO PATIENT ANATOMY OR A FAILURE TO MAINTAIN A STABLE POSITION OF THE DEVICE DURING NEEDLE DEPLOYMENT. NO MANUFACTURING OR QUALITY ISSUE WAS DETECTED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN, TRAINED IN THE USE OF THE PROGLIDE DEVICE, ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED AND A SECOND PROGLIDE WAS USED SUCCESSFULLY TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT EFFECT REPORTED. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 940206H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention