COLLEAGUE TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2011-00132
- Event Type
- Malfunction
- Date Received
- January 7, 2011
- Date of Event
- December 29, 2010
- Report Date
- December 29, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K041191
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF FAILURE (B)(4). THE ROOT CAUSE OF THE REPORTED CONDITION WAS DETERMINED TO BE A LOOSE SPEAKER HARNESS. THE SPEAKER HARNESS WAS RECONNECTED TO REPAIR THE REPORTED CONDITION. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE BAXTER SERVICE TECHNICIAN REPORTED A COLLEAGUE INFUSION PUMP WHICH FAILED TO AUDIBLY ALARM FOR FAILURE CODE 550:320:844:0000 DURING DEVICE EVALUATION. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION NECESSARY. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.03.00. NO FURTHER INFORMATION IS AVAILABLE.