OSCILLATING SAW BLADE 19 X 1.19 X 90MM QTY 5
Report
- Report Number
- 1017294-2024-00070
- Event Type
- Malfunction
- Date Received
- June 13, 2024
- Date of Event
- April 23, 2024
- Report Date
- June 13, 2024
- Manufacturer
- CONMED LARGO
- Product Code
- GFA
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE DEVICE WILL NOT BE RETURNED, AND NO PHOTOGRAPHIC EVIDENCE WAS PROVIDED. THEREFORE, A DEVICE MALFUNCTION CANNOT BE VERIFIED. A TWO-YEAR LOT HISTORY REVIEW CANNOT BE CONDUCTED AS A LOT NUMBER WAS NOT PROVIDED. A DEVICE HISTORY RECORD (DHR) REVIEW CANNOT BE CONDUCTED AS A LOT NUMBER WAS NOT PROVIDED. (B)(4). PER THE INSTRUCTIONS FOR USE, THE USER IS ADVISED THE FOLLOWING: WHEN USING THE POWERPRO SAGITTAL SAW ATTACHMENT (PRO2043), PLACING EXCESSIVE BENDING OR TWISTING FORCE ON THE SAGITTAL SAW BLADE MAY CAUSE THE COLLET TO OPEN AND RELEASE THE SAW BLADE. DO NOT USE SAW BLADES TO PRY, REMOVE BONE GRAFTS, OR AS A LEVERAGE POINT. PATIENT OR USER INJURY COULD OCCUR. WHEN OPERATING THE POWERPRO ELECTRIC II OSCILLATOR HANDPIECE, LET THE SAW BLADE DO THE CUTTING. TOO MUCH FORCE WILL BIND THE BLADE WHICH CAN DAMAGE THE HANDPIECE. WE WILL CONTINUE TO MONITOR FOR TRENDS THROUGH THE COMPLAINT SYSTEM TO ASSURE PATIENT SAFETY.
THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE V190-119-90, OSCILLATING SAW BLADE 19 X 1.19 X 90MM QTY 5 WAS BEING USED ON APPROXIMATELY (B)(6) 2024 DURING A TOTAL KNEE AND ¿1X TOOTH FRONT RIGHT SNAPPED OFF¿. THERE WAS NO REPORT OF IMPACT OR INJURY TO THE PATIENT. AFTER FURTHER ASSESSMENT IT WAS FOUND THAT THE TOOTH WAS RETRIEVED AND DISCARDED. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 938058 | OSCILLATING SAW BLADE 19 X 1.19 X 90MM QTY 5 | BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL | GFA | CONMED LARGO | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |