FDA Adverse Event Malfunction Summary report: N

COLLEAGUE SINGLE CHANNEL CXEPUMP V1.7

MDR report key: 1953024 · Received January 7, 2011

Report

Report Number
6000001-2011-00122
Event Type
Malfunction
Date Received
January 7, 2011
Date of Event
December 1, 2010
Report Date
December 13, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH FOUR UPSTREAM OCCLUSION ALARMS WAS NOT CONFIRMED NOR REPRODUCED DURING PRODUCT EVALUATION BY BAXTER QUALITY ENGINEERING. THEREFORE, NO ASSIGNABLE CAUSE COULD BE PROVIDED FOR THE REPORTED CONDITION AND NO REPAIR WAS NECESSARY. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). PER THE CUSTOMER, THE DEVICE IS AVAILABLE FOR EVALUATION; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE OR ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.

Description of Event or Problem · 1

THE FACILITY REPORTED TO BAXTER (B)(4) THAT A COLLEAGUE INFUSION PUMP EXPERIENCED FOUR UPSTREAM OCCLUSION ALARMS. THESE ALARMS MAY HAVE BEEN FALSE OCCLUSION ALARMS. IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 7.01.00.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE SINGLE CHANNEL CXEPUMP V1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1