Narratives
(B)(4). DEVICE EVALUATION EXPECTED, BUT NOT YET COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF BAXTER'S INVESTIGATION.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE: INSUFFICIENT DRAIN, USE ERROR: TIDAL UF REMOVAL SET TOO LOW. A LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH ((B)(4)).
AN INSTANCE OF INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED DURING A REVIEW OF THE RETURNED HOMECHOICE (HC) PATIENT EVENT LOG. ON (B)(6) 2010, THE DRAIN VOLUME WAS 700ML DURING CYCLE 8.