ECHELON*FLEX60 ARTICULATING
Report
- Report Number
- 1527736-2011-00011
- Event Type
- Malfunction
- Date Received
- January 7, 2011
- Date of Event
- December 13, 2010
- Report Date
- December 14, 2010
- Manufacturer
- ETHICON ENDO SURGERY, INC (CINCINNATI)
- Product Code
- GDW
- PMA / PMN Number
- K081146
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH ONE ECR60B CARTRIDGE RELOAD PRESENT. THE CARTRIDGE RELOAD WAS RECEIVED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. THE RETURN AND TEST CARTRIDGE WERE LOADED INTO THE DEVICE WITHOUT DIFFICULTIES AND DID NOT FELL OUT DURING THE VISUAL AND FUNCTIONAL TESTING. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. A DVD OF THE PROCEDURE WAS RECEIVED, EVIDENCE FROM THE DVD INDICATING THAT THE CARTRIDGE WAS NOT PROPERLY SEATED.
(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS ADDITIONAL INFORMATION: FIELD QUALITY ENGINEER VIEWED DVD: THE BLUE CARTRIDGE APPEARED TO BE NOT PROPERLY SECURED INTO THE CHANNEL. IMAGES OF THE CARTRIDGE APPEARED TO BE NOT PROPERLY SECURED. IN BOTH CASES ABOVE THERE APPEARED TO BE TO MUCH BLUE PLASTIC ABOVE THE CHANNELS EDGE TO BE PROPERLY SECURED. (NOTE THAT THE ANGLE AND MAGNIFICATION CAN DISTORT WHAT APPEARS IN AN IMAGE). AFTER THE DEVICE WAS FIRED AND REMOVAL WAS ATTEMPTED IT APPEARED AS IF THE DISTAL TIP OF THE CARTRIDGE WAS STUCK ON SOMETHING. IT SHOULD BE NOTED THAT THE STAPLE LINES LOOK GOOD WITH GOOD HEMOSTASIS, HENCE THE CARTRIDGE MAY HAVE BEEN DISLODGED AFTER FIRING BY WHAT EVER WAS GRIPPING THE DISTAL TIP. IT SHOULD ALSO BE NOTED THAT THE CARTRIDGE HAD TO BE PULLED OFF WHAT EVER IT WAS ATTACHED TO.
IT WAS REPORTED THAT DURING A LAPAROTOMY PROCEDURE, WHEN THE JAW WAS OPENED AFTER THE FIRING, THE CARTRIDGE CAME OFF FROM THE JAW. ANOTHER COMPETITIVE DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE DEVICE WAS USED ON THE RECTUM DURING THE LAPAROSCOPIC LOW ANTERIOR RESECTION. THE EVENT OCCURRED AFTER THE 1ST FIRING. THE TISSUE WAS NOT SO THICK. THE DEVICE WAS NOT FIRED ACROSS OR NEAR THE HARD OBJECTS. REINFORCEMENT MATERIAL WAS NOT USED. WHEN THE CARTRIDGE WAS LOADED INTO THE JAW, THE SNAP SOUND WAS NOT CONFIRMED. THE DEVICE HAD NOT DIFFICULTY IN CLOSING BUT WE HAVE NO DETAILED INFORMATION WHETHER THERE WAS DIFFICULTY IN FIRING OR NOT. THE STAPLES WERE DEPLOYED ON THE TISSUE AND THEY LOOKED LIKE NO PROBLEMS. THE DOCTOR COMMENTED THAT THE JAW SEEMED TO TOUCH THE PUBIS WALL DURING THE OPERATION. THE FALLEN CARTRIDGE WAS RETRIEVED BY A FORCEPS. THE PATIENT IS GETTING WELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ECHELON*FLEX60 ARTICULATING | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO SURGERY, INC (CINCINNATI) | UNK | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RELOAD - (B)(4) |