FDA Adverse Event Malfunction Summary report: N

VITROS ECIQ IMMUNODIAGNOSTIC SYSTEM

MDR report key: 1952872 · Received January 7, 2011

Report

Report Number
1319681-2011-00005
Event Type
Malfunction
Date Received
January 7, 2011
Date of Event
December 9, 2010
Report Date
January 7, 2011
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
MMI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION DETERMINED THAT A NON-REPRODUCIBLE HIGHER THAN EXPECTED VITROS TROP I ES RESULT WAS OBTAINED FROM A SINGLE PATIENT SAMPLE PROCESSED ON THE VITROS ECIQ IMMUNODIAGNOSTIC SYSTEM. AN OCD FIELD ENGINEER REPLACED THE WELL WASH DISPENSE VALVE AND CALIBRATED WELL WASH DISPENSE VOLUMES. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, THE EVENT IS MOST LIKELY INSTRUMENT RELATED.

Description of Event or Problem · 1

THE CUSTOMER OBTAINED A NON-REPRODUCIBLE HIGHER THAN EXPECTED VITROS TROP I ES RESULT FROM A SINGLE PATIENT SAMPLE PROCESSED ON THE VITROS ECIQ IMMUNODIAGNOSTIC SYSTEM. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION. THE HIGHER THAN EXPECTED RESULT WAS REPORTED OUT OF THE LABORATORY. A CORRECTED REPORT WAS ISSUED FOR THE AFFECTED PATIENT SAMPLE. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B)(4)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITROS ECIQ IMMUNODIAGNOSTIC SYSTEM IMMUNODIAGNOSTIC ANALYZER MMI ORTHO-CLINICAL DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1