VITROS ECIQ IMMUNODIAGNOSTIC SYSTEM
Report
- Report Number
- 1319681-2011-00005
- Event Type
- Malfunction
- Date Received
- January 7, 2011
- Date of Event
- December 9, 2010
- Report Date
- January 7, 2011
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- MMI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION DETERMINED THAT A NON-REPRODUCIBLE HIGHER THAN EXPECTED VITROS TROP I ES RESULT WAS OBTAINED FROM A SINGLE PATIENT SAMPLE PROCESSED ON THE VITROS ECIQ IMMUNODIAGNOSTIC SYSTEM. AN OCD FIELD ENGINEER REPLACED THE WELL WASH DISPENSE VALVE AND CALIBRATED WELL WASH DISPENSE VOLUMES. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, THE EVENT IS MOST LIKELY INSTRUMENT RELATED.
THE CUSTOMER OBTAINED A NON-REPRODUCIBLE HIGHER THAN EXPECTED VITROS TROP I ES RESULT FROM A SINGLE PATIENT SAMPLE PROCESSED ON THE VITROS ECIQ IMMUNODIAGNOSTIC SYSTEM. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION. THE HIGHER THAN EXPECTED RESULT WAS REPORTED OUT OF THE LABORATORY. A CORRECTED REPORT WAS ISSUED FOR THE AFFECTED PATIENT SAMPLE. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B)(4)