FDA Adverse Event Injury Summary report: N

PTERYFIT IMPLANT

MDR report key: 19527186 · Received June 13, 2024

Report

Report Number
3011390931-2024-00296
Event Type
Injury
Date Received
June 13, 2024
Date of Event
April 2, 2024
Report Date
June 13, 2024
Manufacturer
NORIS MEDICAL LTD
Product Code
DZE
UDI-DI
07290108817102
PMA / PMN Number
K140440
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DHR WAS REVIEWED FOR LOT#9000920 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE INVESTIGATION FOR THIS EVENT HAS NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. OUR MANUFACTURING SYSTEM ASSURES, THAT PRODUCTION AND PROCESS CONTROLS ARE IN PLACE TO ENSURE THAT BATCHES CONFORM TO THE APPLICABLE SPECIFICATIONS BEFORE THEY ARE DISTRIBUTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

UDI RELATED DATA QUALITY UPDATES ONLY.

Description of Event or Problem · 0

A HEALTHCARE PROFESSIONAL REPORTED THAT NMAF4218 LOT#9000920 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
864142 PTERYFIT IMPLANT PTERYFIT DENTAL IMPLANT DZE NORIS MEDICAL LTD NMAF4218 9000920 07290108817102

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown