FDA Adverse Event Malfunction Summary report: N

DELTA

MDR report key: 1952702 · Received January 4, 2011

Report

Report Number
1220063-2010-00097
Event Type
Malfunction
Date Received
January 4, 2011
Date of Event
November 25, 2010
Report Date
December 2, 2010
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC.
Product Code
MHX
PMA / PMN Number
UNK
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER'S A105 PROCESSOR PCBA WAS EVALUATED AND FOUND TO HAVE A BURNT CAPACITOR IN THE BATTERY CHARGING CIRCUIT. THERE WERE BURN MARKS BUT NO PCB DAMAGE. A CONTINUITY CHECK WAS DONE TO VERIFY THAT THE REMAINING PARALLEL CAPACITORS IN THE CIRCUIT WERE NOT SHORTED. THE BURNT CAPACITOR AND TWO ADDITIONAL CAPACITORS WERE SENT TO THE SUPPLIER FOR FURTHER ANALYSIS, BUT THE SUPPLIER WAS UNABLE TO DETERMINE THE EXACT CAUSE FOR THE FAILED CAPACITOR. THERE WERE NO MFG PROBLEMS NOTED FROM THE MANUFACTURE DATE OF THESE CAPACITORS. THE ROOT CAUSE FOR THE FAILED CAPACITOR WITHIN THE PROCESSOR BOARD OF THE MONITOR IS UNK. DRAEGER PERFORMED A HEALTH HAZARD EVAL AND DETERMINED THAT THE STATISTICAL RISK FOR A SIMILAR OCCURRENCE IS CONSIDERED TO BE REMOTE. DRAEGER WILL CONTINUE TO MONITOR FOR SIMILAR REPORTS OF THIS CONDITION.

Description of Event or Problem · 1

IT WAS REPORTED THAT A NURSE NOTICED A SMELL COMING FROM A DELTA MONITOR AS IF SOMETHING HAD BURNT. IT STARTED TO ALARM AND THEN SHUT DOWN. THERE WAS NO PT INJURY REPORTED. DRAEGER REFERENCE NUMBER: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DELTA PATIENT MONITOR MHX DRAEGER MEDICAL SYSTEMS, INC. UNK NA

Patients

Seq Age Sex Outcome Treatment
1 UNK NO